HU5-S2

Pendant Service System (Ceiling Mounted)

Description

The HU5-S2 unit suspends equipment within the patient area and is a single, compact unit.

Areas of Use:

High Dependency, Operating Theatres

 

Design:

The HU5-S2 is a compact single ceiling pendant solution for suspending equipment within the patient area;

The design comprises a single, dual arm articulated vertical service column;

The adjustable friction brake articulated arm and service column can be rotated within 320 degrees to position the Supply Services ergonomically;

The design provides unobstructed access around the patient area;

The service column is populated to suit specific needs and the centre fascia can be customised to accommodate interior design intent;

4 x Ø38mm equipment poles are integrated on each corner of the service column to support medical equipment (*see Attachment range to suit the medical equipment);

Self-clamping corner extrusions secure the fascia eliminating visible screws;

As a standard, the fascias are offered in a powder coated finish (RAL 9002) with other options available on request.

 

Services:

The moveable dual articulated arm service column provides for low voltage (LV), extra low voltage (ELV) and selected medical gases (MG);

The product can accommodate various types of electrical and medical gas outlets, fitted to suit the clients’ specifications;

All four sides of the service column are populated with Supply Services to support medical equipment;

Anodized aluminium construction provides a rigid equipment platform.

 

Payload:

Dual articulated arm service column: <60kg

 

Installation:

The unit is flat slab mounted*, where Supply Services are terminated in the ceiling void;

The unit is manufactured with prepopulated electrical and medical gas outlets, which are tested and validated against international standards, prior to installation;

The unit is supplied with mounting structure and specialised anchors to suit site requirements;

Installation to be provided by qualified Hutz Medical installation engineers;

(*other mounting options are available on request).

 

Norms & Certifications:

IEC 60601-1

IEC 60601-2

ISO 11197